Fluke 1587 Insulation Multimeters
Specifications |
Voltage DC |
Maximum Voltage |
1000 V |
Accuracy |
±(0.09%+2) |
Maximum Resolution |
0.001V |
Voltage AC |
Maximum Voltage |
1000 V |
Accuracy |
±(2%+3) |
AC Bandwidth |
5kHz* *with low pass filter; 3db @ 800 Hz |
Maximum Resolution |
0.1 mV |
Current DC |
Maximum Amps |
400mA |
Amps Accuracy |
±(0.2%+2) |
Maximum Resolution |
0.01 mA |
Current AC |
Maximum Amps |
400mA |
Amps Accuracy |
±(1.5%+2)* * 1kHz bandwidth |
Maximum Resolution |
0.01 mA |
Resistance |
Maximum Resistance |
50 MΩ |
Accuracy |
±(0.9%+2) |
Maximum Resolution |
0.1Ω |
Capacitance |
Maximum Capacitance |
9,999 µF |
Accuracy |
±(1.2%+2) |
Maximum Resolution |
1 nF |
Frequency |
Maximum Frequency |
100 kHz |
Accuracy |
±(0.1%+1) |
Maximum Resolution |
0.01 Hz |
Temperature Measurement |
-40.0 °C to 537 °C -40.0 °F to 998 °F excluding probe |
Diode Test |
Range |
6 V |
Resolution |
1 mV |
Accuracy |
±(2 % + 1) |
Insulation Test |
Minimum Test Current @ 1kΩ/V |
1mA |
Test Voltages |
50, 100, 250, 500, 1000 V |
Maximum Resistance per Test Voltage |
50V |
|
100V |
|
250V |
|
500V |
|
1000V |
2 GΩ |
Maximum Resolution per Test Voltage |
50V |
0.01 MΩ |
100V |
0.01 MΩ |
250V |
0.1 MΩ |
500V |
0.1 MΩ |
1000V |
0.1 MΩ |
Accuracy per Test Voltage |
50V |
±(3 % + 5) |
100V |
±(3 % + 5) |
250V |
±(1.5 % + 5) |
500V |
±(1.5 % + 5) |
1000V |
±(1.5 % + 5) to 600 MΩ, ±(10 % + 3) above 600 MΩ |
Environmental Specifications |
Operating temperature |
–20 °C to +55 °C |
Storage temperature |
-40°C to +60°C |
Humidity (Without Condensation) |
0% - 95% (10 °C – 30 °C) |
0% - 75% (30 °C – 40 °C) |
0% - 40% (40 °C – 55 °C) |
Operating Altitude |
2000 m |
Safety Specifications |
Overvoltage category |
Complies with ANSI/ISA 82.02.01 (61010-1) 2004, CAN/CSA-C22.2 NO. 61010-1-04, and IEC/EN 61010-1 2nd Edition for measurement CAT III 1000 V and CAT IV 600 V |
Agency Approvals |
2000 m CAT III 1000 V, CAT IV 600 V; 3000 m CAT II 1000 V, CAT III 600 V |
Mechanical & General Specifications |
Size |
203 x 100 x 50 mm (with holster) |
Weight |
624 g |
Warranty |
3 years |
Battery life Alkaline |
Meter use |
1000 hours |
Insulation test use |
Meter can perform at least 1000 insulation tests with fresh alkaline batteries at room temperature. These are standard tests of 1000 V into 1 MΩ with a duty cycle of 5 seconds on and 25 seconds off. |
IP Rating |
IP40 |
i400 AC Current Clamp
Specifications |
Nominal current range |
400A |
Continuous current range |
1 A - 400 A |
Maximum Non-Destructive Current |
1000 A |
Lowest measurable current |
1 A |
Basic accuracy |
2% + 0.06 A (45-400 Hz) (% reading + floorspec) |
Useable frequency |
5 Hz - 20 kHz |
Output level(s) |
1 mA/A |
Safety Specifications |
Safety |
CAT IV 600 V, CAT III, 1000 V |
Maximum voltage |
1000 V AC rms from earth to ground, in compliance with EN61010 |
Mechanical & General Specifications |
Warranty |
1 year |
Maximum conductor diameter |
32 mm (1.25 inches) |
Maximum conductor size (busbar) |
750 MCM |
Output cable length |
1.5 m (59 inches) |
Shrouded banana plugs |
Yes |
Fluke 62 MAX+ Infrared Thermometer
Specifications |
Temperature Range
|
-30 °C to 650 °C (-22 °F to 1202 °F) |
Accuracy |
±1.0 °C or ±1.0% of reading, whichever is greater -10 °C to 0 °C: ±2.0 -30 °C to -10 °C: ±3.0 |
Response Time (95%) |
Fluke 62 MAX+ |
<300 ms (95% of reading) |
Spectral Response |
8 to 14 microns |
Emissivity |
0.10 to 1.00 |
Optical Resolution
|
12:1 (calculated at 90% energy) |
Display Resolution |
0.1 °C (0.2 °F) |
Repeatability of Readings |
±0.5% of reading or <±0.5 °C (1 °F), whichever is greater |
Power |
AA battery |
Battery Life |
8 hours with laser and backlight on |
Physical Specifications |
Weight |
255 g (8.99 oz) |
Size |
(175 x 85 x 75) mm (6.88 x 3.34 x 2.95) inches |
Operating Temperature |
0 °C to 50 °C (32 °F to 122 °F) |
Storage Temperature |
-20 °C to 60 °C (-4 °F to 140 °F), (without battery) |
Operating Humidity |
10 % to 90 % RH non-condensing at 30 °C (86 °F) |
Operating Altitude |
2000 meters above mean sea level |
Storage Altitude |
12,000 meters above mean sea level |
IP Rating |
IP 54 per IEC 60529 |
Drop Test |
3 meters |
Vibration and Shock |
IEC 68-2-6 2.5 g, 10 to 200 Hz, IEC 68-2-27, 50 g, 11 ms |
EMC |
EN 61326-1:2006 EN 61326-2:2006 |
Standards and Agency Approval
|
Compliance |
EN/IEC 61010-1: 2001 |
Laser Safety |
FDA and EN 60825-1 Class II |
|
|
|
|